![Agilent Technologies 4294A Operating Manual Download Page 337](http://html.mh-extra.com/html/agilent-technologies/4294a/4294a_operating-manual_2867884337.webp)
Chapter 10
329
Specifications and Supplemental Performance Characteristics
Basic Characteristics
10.
Sp
ec
ifi
ca
tio
ns
and
S
upp
le
men
tal Per
fo
rm
anc
e
C
har
ac
ter
isti
cs
Measurement Time
Figure 10-2
Measurement Time (SPC)
NOTE
When the 42941A Impedance Probe or the 42942A Terminal Adapter is used,
measurement time is 1.5 times longer than the value shown in Figure 10-2.
Trigger Function
Measurement Bandwidth/Averaging
Trigger types
Continuous, single, number of groups
Trigger sources
Internal (free run), external (BNC connector input),
GP-IB or LAN, manual (front key)
Trigger event types
Point trigger, sweep trigger
Measurement bandwidth
Range
1 (Fast) to 5 (Precise), 5 steps
Averaging
Types
Sweep-to-sweep averaging, point averaging