232
Chapter 8
Analysis and Processing of Result
Define the Peak
Define the Peak
When peaks are defined, unwanted peaks can be excluded so that only the peaks matching
the definition are detected.
In Figure 8-14, the Agilent 4294A peaks are defined.
Figure 8-14
Defining the Peak
Definition of peak polarity
Define the peak polarity according to the following steps:
Step 1.
Activate the trace for which the peak polarity is defined.
Step 2.
Press the
[Search]
key to display the Search menu.
Step 3.
Press the
PEAK
key to display the Peak menu.
Step 4.
Press the
PEAK DEF MENU
key to display the Peak Definition menu.
Step 5.
Press the
PEAK PLRTY [ ]
key to select the peak polarity (pressing the key toggles the
peak polarity).
Active Trace
Key operation
Trace A
[A]
Trace B
[B]
Softkey label
Selected peak polarity
PEAK PLRTY [POS]
Positive (if the measurement at this point is
greater than the measurement of the adjacent
measuring points, it is regarded as a peak)
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......