324
Chapter 10
Specifications and Supplemental Performance Characteristics
Basic Characteristics
Specifications describe the instrument's warranted performance over the temperature range
of 0
°
C to 40
°
C (except as noted). Supplemental performance characteristics are intended to
provide information that is useful for applying non-warranted performance parameters to
the instrument. Such parameters are denoted as SPC (supplemental performance
characteristics), typical, or nominal. Warm-up time must at least 30 minutes after power-on
for all specifications.
Basic Characteristics
Measurement Parameter
Measurement Terminal
Source Characteristics
Frequency
Voltage Signal Level
Impedance Parameters
|Z|-
q
, R-X, Ls-Rs, Ls-Q, Cs-Rs, Cs-Q, Cs-D, |Y|-
q
,
G-B, Lp-G, Lp-Q, Cp-G, Cp-Q, Cp-D, Complex Z-Y,
|Z|-Ls, |Z|-Cs, |Z|-Lp, |Z|-Cp, |Z|-Rs, |Z|-Q, |Z|-D,
Lp-Rp, Cp-Rp
Configuration
Four-terminal pair configuration
Connector type
Four BNC (female) connectors. Can be converted to
one-port terminal using the 42942A Terminal Adapter
(7-mm port) or the 42941A Impedance Probe
(3.5-mm port).
Range
40 Hz to 110 MHz
Resolution
1 mHz
Accuracy
without Option 1D5
± 20 ppm (at 23 ± 5
°
C)
± 40 ppm (at 0 to 55
°
C)
with Option 1D5
± 0.13 ppm (at 0 to 55
°
C)
Range
5 mVrms to 1 Vrms
Resolution
1 mV
Accuracy
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......