![Agilent Technologies 4294A Operating Manual Download Page 162](http://html.mh-extra.com/html/agilent-technologies/4294a/4294a_operating-manual_2867884162.webp)
154
Chapter 6
Calibration
Fixture Compensation
Step 4.
Select one of the compensation data definition keys,
OPEN CONDUCT(G)
,
OPEN CAP(C)
,
SHORT RESIST(R)
,
SHORT INDUCT(L)
,
LOAD RESIST(R)
, or
LOAD INDUCT(L)
, to
confirm or set the compensation data definition.
The current value of the compensation data definition is displayed in the setting parameter
area of the LCD's upper-left side.
Step 5.
Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the compensation data definition .
•
Enter the desired value with the numerical keys (
[0]
through
[9]
,
[.]
,
[-]
) and then press
one of the unit keys (
[G/n]
,
[M/
m
]
,
[k/
m
]
, or
[
´
1]
).
•
Turn the rotary knob (
m
m
m
m
) until the desired value is set.
•
Press the step keys (
[
][
¯
]
) to set the desired value.
NOTE
When entering the capacitance value in the compensation data definition, the
[fF]
unit is
automatically added at the end of the value you enter with the unit keys. For example, if
you enter a value with the numeric keys and then press the
[x1]
key, the
[fF]
unit is added to
the entered value.
Step 6.
Repeat steps 4 and 5 to set all of the required compensation data definitions.
Compensation Data Definition
Key Stroke
Equivalent parallel conductance value (G) of
OPEN standard used for fixture compensation
(preset state: 0 S)
OPEN CONDUCT(G)
Equivalent parallel capacitance value (C) of
OPEN standard used for fixture compensation
(preset state: 0 F)
OPEN CAP(C)
Equivalent series resistance value (R) of
SHORT standard used for fixture
compensation
(preset state: 0
W
)
SHORT RESIST(R)
Equivalent series inductance value (L) of
SHORT standard used for fixture
compensation
(preset state: 0 H)
SHORT INDUCT(L)
Equivalent series resistance value (R) of
LOAD standard used for fixture compensation
(preset state: 50
W
)
LOAD RESIST(R)
Equivalent series inductance value (L) of
LOAD standard used for fixture compensation
(preset state: 0 H)
LOAD INDUCT(L)
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......