![Agilent Technologies 4294A Operating Manual Download Page 334](http://html.mh-extra.com/html/agilent-technologies/4294a/4294a_operating-manual_2867884334.webp)
326
Chapter 10
Specifications and Supplemental Performance Characteristics
Basic Characteristics
Signal Level Monitor
NOTE
f
: frequency [MHz],
Zx
: impedance measurement value [W].
Beyond 23 ± 5 ×C, the test signal level monitor accuracy decreases to half the value
described.
Output Impedance
dc Bias Function
Voltage range
(same as the voltage signal level setting range)
Voltage monitor accuracy
at four-terminal pair port of
the 4294A or 7-mm port of
the 42942A
± (10 + 0.05
´
f
+ 100/
Z
x
) [%]
(at 23 ± 5
°
C, typical)
at measurement port of the
42941A, 16048G/H
± (10 + 0.15
´
f
+ 100/
Z
x
) [%]
(at 23 ± 5
°
C, typical)
Current range
(same as the current signal level setting range)
Current monitor accuracy
at four-terminal pair port of
the 4294A or 7-mm port of
the 42942A
± (10 + 0.3
´
f
+
Z
x
/100) [%]
(at 23 ± 5
°
C, typical)
at measurement port of the
42941A, 16048G/H
± (10 + 0.4
´
f
+
Z
x
/100) [%]
(at 23 ± 5
°
C, typical)
Output impedance
25
W
(nominal)
dc voltage bias
Range
0 to
±
40 V (see Figure 10-1)
Resolution
1 mV
Accuracy
±
[0.1% + (5 + 30
´
|
I
mon
|) mV] (at 23
±
5
°
C)
±
[0.2% + (10 + 30
´
|
I
mon
|) mV] (beyond 23
±
5
°
C)
dc current bias
Range
0 to
±
100 mA (see Figure 10-1)
Resolution
40
m
A
Accuracy
±
[2% + (0.2 + |
V
mon
|/20) mA] (at 23
±
5
°
C)
±
[4% + (0.4 + |
V
mon
|/10) mA] (beyond 23
±
5
°
C)
dc voltage bias at constant
voltage mode
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......