18-11
IM 701240-01E
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
18
App
Index
18.11 General Specifications
Item
Specifications
Standard operating
Ambient temperature:
23
±
5
°
C
conditions
Ambient humidity:
20 to 80%RH
Within 1% of the rated error of the power supply voltage and frequency
After a 30-minute warm-up and after calibration
Recommended calibration 1 year
period
Warm-up time
At least 30 minutes
Storage temperature
–20 to 60
°
C
Storage humidity
20 to 85%RH (no condensation)
Storage altitude
3000 m or less
Operating temperature
5 to 40
°
C
range
Operating humidity range
20 to 85%RH (when not using the printer), 35 to 85% RH (when using the printer)
Operating altitude
2000 m or less
Rated supply voltage
100 to 120 VAC or 200 to 240 VAC (automatic switching)
Rated power supply
50/60 Hz
frequency
Permitted supply voltage
90-132 VAC/180-264 VAC
Permitted power supply
48 to 63 Hz
frequency range
Maximum power
Approx. 200 VA max. (maximum power when the printer is OFF and 16 channels are running is 135
consumption
VA (reference value))
Withstand voltage
1500 VAC for 1 minute across the power supply and earth
Insulation resistance
10 M
Ω
or more at 500 VDC across the power supply and ground
External dimensions
355 mm (W)
×
250 mm (H)
×
180 mm(D) (excluding the handle and other projections)
Weight
Approx. 7.8 kg (only the SL1400 with all options (C8/C10/P4 options), without the chart paper)
Approx. 11.8 kg ( eight High-Speed 10 MS/s, 12-Bit Isolation Modules)
Module: Approx. 300 g (High-Speed 10 MS/s, 12-Bit Isolation Module)
Instrument’s cooling
Forced air cooling. Exhaust on the left side panel and top.
method
Battery backup
Settings and clock are backed up with the internal lithium battery
Battery backup
Approx. 5 years (at ambient temperature of 25
°
C)
Battery life
Fuse