17-7
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
17
18
App
Index
Troubleshooting, Maintenance, and Inspection
Code
Message and Corrective Action
Section
932
Slot 3 EEPROM error.
–
933
Slot 4 EEPROM error.
–
934
Slot 5 EEPROM error.
–
935
Slot 6 EEPROM error.
–
936
Slot 7 EEPROM error.
–
937
Slot 8 EEPROM error.
–
938
Key protect is enabled. To release the protection, press the PROTECT key.
16.5
940
The USB device’s power consumption exceeded the capacity of the USB hub.
18.8
File Operation Errors (1000 to 1199)
Code
Message and Corrective Action
Section
1001
Invalid file name.
Chapter 13
* Duplicate file name or incorrect SCSI ID.
1002
Cannot detect the medium.
Chapter 13
* Check the presence of the medium or the SCSI device connection.
1003
Cannot detect the medium.
Chapter 13
* Check the presence of the medium or the SCSI device connection.
1004
Media failure.
Chapter 13
Check the storage medium.
1005
File not found.
Chapter 13
Check the file name and the storage medium.
1006
Invalid file name.
Chapter 13
* Duplicate file name or incorrect SCSI ID.
1007
Media failure.
Chapter 13
Check the storage medium.
1008
Invalid file name.
Chapter 13
* Duplicate file name or incorrect SCSI ID.
1009
Invalid file name.
Chapter 13
* Duplicate file name or incorrect SCSI ID.
1010
Invalid file name.
Chapter 13
* Duplicate file name or incorrect SCSI ID.
1011
The maximum number of files that can be stored in a single directory was exceeded. Save the
Chapter 13
file to another directory or medium.
1012
Media full. Delete unnecessary file(s) or use another disk.
Chapter 13
1013
Cannot delete a directory if there are files in the directory.
13.16
1014
File is protected.
13.16
1015
Physical format error. Reformat the medium.
13.6
If the same error occurs, the instrument is probably unable to execute a format on this medium.
1016
File system failure.
Chapter 13
1017
File system failure. Check using another disk.
Chapter 13
If the same message still appears, maintenance service is required.
1018
File system failure. Check using another disk.
Chapter 13
If the same message still appears, maintenance service is required.
1019
File system failure. Check using another disk.
Chapter 13
If the same message still appears, maintenance service is required.
1020
File system failure. Check using another disk.
Chapter 13
If the same message still appears, maintenance service is required.
1021
File is damaged. Check the file.
Chapter 13
1022
File system failure. Check using another disk.
Chapter 13
If the same message still appears, maintenance service is required.
1023
File system failure. Check using another disk.
Chapter 13
If the same message still appears, maintenance service is required.
17.2 Messages and Corrective Actions