13-15
IM 701240-01E
Saving and Loading Data
3
2
1
4
5
6
7
8
9
10
11
12
13
13
14
15
16
17
18
App
Index
Selecting the Range of the Waveform to Be Saved
9.
Press the
Range
soft key. A menu used to select the range of the waveform to
be saved appears.
10.
Press the
Main
,
Z1
, or
Z2
soft key to select the range of the waveform to be
saved.
11.
Press the
History
soft key to save all the data in the history memory (All), save
only the selected waveform (One), or save the average data in the history
memory (Ave).
If you select All after performing a search on the history memory data, only the
searched waveforms are saved.
Compressing or Decimating the Data and Saving the Data
Decimating the Data and Saving (When Data Type Is Set to ASCII)
12.
Press the
Save Interval
soft key.
Turn the
jog shuttle
to set the save interval from OFF, Per5 to Per5000.
Compressing the Data and Saving (When Data Type Is Set to Binary)
12.
Press the
P-P Comp
soft key to select ON (compress and save) or OFF (do not
compress and save).
If P-P Comp is turned ON when saving waveform data, only the maximum and
minimum values of the multiple data points existing at the same time position
are saved. Consequently, the file size is reduced. However, the SL1400 cannot
load the saved data.
Note
If P-P Comp is set to ON, you cannot set the save range using Range.
For Data
Type: Binary
For Data
Type: ASCII
13.8 Saving and Loading Waveform Data