9-9
IM 701240-01E
Recording in Recorder Mode
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Dark/Light
Set the graticule dark/light setting to Light or Dark.
Width of the Vertical Graticule (Width)
Select how to set the vertical graticule width.
1Div:
Grid that divides the recording area into 10 areas
10mm:
mm graticule type
Grid Type
Graticule
Width
1 Div
10 mm
1 div
10 mm
1 div
10 mm
No graticule lines
No graticule lines
OFF
Graticule lines that are printed
The format of the scale lines that are printed depending on the selected graticule
type and width is given in the table below.
Details
Print Interval of Scale Values (Scale)
Select the interval for the scale values (see section 8.9) printed at the top and bottom
edges of the recording area from OFF, 200 mm, 400 mm, or 800 mm. If OFF is
selected, the scale values are not printed.
Print Interval of Channel Labels (Ch Label)
Select the interval for printing waveform channel labels (see section 8.10) from OFF,
200 mm, 400 mm, or 800 mm. The channel label is printed near each waveform. If
OFF is selected, the channel labels are not printed.
Time Print
Select the interval for printing the time from OFF, 200 mm, 400 mm, or 800 mm. The
time is printed at the top row of the recording area. If OFF is selected, the times are
not printed.
Gauge Print
Select whether to not print (OFF), print (ON), or print in detail (Fine) on the right (or
left) side of the recording area. If ON is selected a gauge that equally divides the
vertical axis of the waveform display area into two is displayed. If Fine is selected, a
gauge that equally divides the vertical axis of the waveform display area into 10 is
displayed. In realtime recording, the gauge is printed on the right side of the recording
area when the recording is completed. When reprinting waveforms saved to the
internal memory (see section 9.7), the gauge is printed on the left side of the
recording area when the recording is started. The gauge, scale values, an arrow
indicating the ground position, and measurement range* are printed.
*
The measurement range may not be printed depending on the setting of T-Y Recording.
Paper Feed after Stopping the Recording
You can select whether the paper is fed automatically after stopping the T-Y
waveform recording. This function can be used on products with firmware version 6.52
or higher. Check the firmware version under Soft Version described in section 17.4.
ON:
Automatically feeds the paper approximately 40 mm so that the gauge
information that is printed after the waveform recording can be seen.
OFF:
Disables the auto feed function. To feed the paper, do it manually
according to the procedure given on page 3-12.
9.3 Setting the Recording Format of T-Y Waveform Recording