Triggering
Explanation
This function activates a trigger when the trigger source signal crosses a specified level.
Trigger Source
Select the trigger source from CH1 to CH16.
Note
If you set the trigger source to a value other than CH1 to CH16, see the sections below.
•
Time:
Section 6.8, “Setting the Timer Trigger (SIMPLE)”
•
LogicA or Logic B:
Section 6.9, “Setting the Logic Trigger (SIMPLE)”
•
Ext (TRIG IN terminal of the side panel): Section 6.6, “Setting the External Trigger (SIMPLE)”
•
Line (commercial power supply signal):
Section 6.7, “Generating Triggers on the Power
Signal (SIMPLE)”
Trigger Level
When Measuring Voltage
Selectable range:
Within voltage range
Resolution:
Varies depending on the input module (see the table below).
When Measuring Temperature
Selectable range:
Measurement range of each thermocouple
Resolution:
0.1
°
C or 0.1 K (set within the measurement range (varies
depending on the thermocouple type)
For a description of the measurement range of thermocouples,
see section 5.14.
When Measuring Strain
Selectable range:
±
(Measurable range)
Resolution:
1
µ
STR or 0.0005 mV/V
Trigger Level Resolution
The selectable resolution varies depending on the input module.
When Measuring Voltage (When Probe Attenuation Is 1:1)
Input Module
701250/701255/
701251
701265
701267
701275
Voltage Range
701261/701262
1 mV
–
–
0.001 mV
–
–
2mV
–
–
0.002 mV
–
–
5 mV
–
–
0.005 mV
–
–
10 mV
–
0.01 mV
0.01 mV
–
–
20 mV
–
0.02 mV
0.02 mV
–
–
50 mV
0.05 mV
0.05 mV
0.05 mV
–
0.05 mV
100 mV
0.1 mV
0.1 mV
0.1 mV
–
0.1 mV
200 mV
0.2 mV
0.2 mV
0.2 mV
0.2 mV
0.2 mV
500 mV
0.5 mV
0.5 mV
0.5 mV
0.5 mV
0.5 mV
1 V
0.001 V
0.001 V
0.001 V
0.001 V
0.001 V
2 V
0.002 V
0.002 V
0.002 V
0.002 V
0.002 V
5 V
0.005 V
0.005 V
0.005 V
0.005 V
0.005 V
10 V
0.01 V
0.01 V
0.01 V
0.01 V
0.01 V
20 V
0.02 V
0.02 V
0.02 V
0.02 V
0.02 V
50 V
0.05 V
0.05 V
0.05 V
0.05 V
0.05 V
100 V
0.1 V
0.1 V
0.1 V
0.1 V
0.1 V
200 V
0.2 V
0.2 V
–
0.2 V
–
500 V
–
–
–
0.5 V
–
1 kV
–
–
–
0.001 kV
–
2 kV
–
–
–
0.002 kV
–
The resolution is 0.01 division on all modules.
6.5 Setting the Edge Trigger (SIMPLE)