6-44
IM 701240-01E
Window Width
The realtime template is created by applying a tolerance width (window width) to the
averaged waveform. The selectable range of the window width varies depending on the
measured item as follows:
When measuring voltage:
(Voltage range)/1000 to voltage range
When measuring strain:
11
µ
STR to (measurement range)
×
2
Or, 0.0005 mV/V to (measurement range)
×
2
When measuring acceleration:
0.01 Unit to (Unit/div)
×
10
(The resolution is the same as the resolution of the trigger level. See section 6.5.)
For example, if Width is set to 2 V, a window width of
±
1 V is created with respect to the reference waveform.
Window width
1 V
1 V
Reference waveform
Hold Off
See section 6.4.
Note
If a trigger waveform such as a surge is input immediately before the completion of the pre-
trigger, the trigger point may be delayed by up to 4 cycles.
6.17 Setting the Wave Window Trigger (ENHANCED)