Waveform Acquisition
Notes on Averaging
• Averaging is effective only for repetitive waveforms.
• Correct averaging is not possible if the waveform has imperfect triggering (incomplete
synchronization). Thus, the displayed waveform will be distorted. When working with
this type of signal, set the trigger mode to Normal, so that the waveform display is
updated only when the trigger is activated (see section 6.1).
• Roll mode display is disabled during averaging.
• Averaging is not possible when the trigger mode is Single, Single (N), or Log.
• If you stop the waveform acquisition by pressing START/STOP, the averaging
process also stops. Averaging restarts from the beginning when acquisition resumes.
• If you are using simple averaging, the SL1400 terminates acquisition automatically
when the specified number of acquisitions (acquisition count) is completed.
• Averaging is not possible when realtime recording is in progress.
• When acquiring waveforms in averaging mode, the data that is retained in the history
memory is 1 record.
7.2 Setting the Acquisition Mode