Preparations for Measurements
• For the 701275 (ACCL/VOLT)
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 passive probe (10:1)
11
or direct input (cable that
does not comply with the safety standards)
9
42 V (DC+ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 passive probe (10:1)
12
or direct input (cable that
does not comply with the safety standards)
10
42 V (DC+ACpeak) (CAT II, 30 Vrms)
• For the 701280 (FREQ)
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the isolated probe 700929 (10:1)
1
420 V (DC+ACpeak)
• Safety cable (1:1) (combined with 701954)
5
or direct input (cable that
does not comply with the safety standards)
9
42 V (DC+ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• Combined with the isolated probe 700929 (10:1)
2
300 Vrms (CAT II)
• Safety cable (1:1) (combined with 701954)
8
or direct input (cable that
does not comply with the safety standards)
10
42 V (DC+ACpeak) (CAT II, 30 Vrms)
With the 700929
1
2
700929
With the 701901 and 701954
H
L
Direct input (cable that does not comply
with the safety standards)
9
10
BNC
3
4
With the 10:1 passive probe (701940)
H
L
11
12
H
L
701901
701954
5
H
L
8
6
7
Over-Range Indication
If over-range is indicated, the SL1400 may be receiving a voltage higher than
the observed waveform or measured waveform values. To prevent electric
shock, change the vertical scale with the RANGE knob so that the entire
amplitude of the waveform is displayed within the waveform display area, and
check the input voltage level.
Over-range indication
Indicates the number of the channel that over-range is occurring on.
If over-range is occurring on multiple channels, the smallest channel
number among them is indicated.
3.7 Connecting Probes