6-3
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Triggering
6.2
Setting the Trigger Position
<For a description of this function, refer to page 2-25.>
Procedure
1.
Press
MENU
.
2.
Press the
TRIGGER
soft key.
3.
Press the
Position
soft key.
4.
Turn the
jog shuttle
to set the trigger position. Pressing
RESET
resets the
value to 50%.
Explanation
Trigger Position
Trigger position = Trigger point + trigger delay
You can select the location of the trigger position on the screen.
If the trigger delay is 0 s, the trigger position and the trigger point match.
Delay
Trigger point
T (Trigger position)
Selectable Range of the Trigger Position
With the display record length taken to be 100%, set in the range of 0 to 100% (in 0.1%
steps).
Displaying the Trigger Position
A position marker ( ) appears at the top of the screen. The marker indicates the trigger
position with respect to the display record length.