12-8
IM 701240-01E
Flexible Zone
The flexible zone can be set only when the format is 1 Zone and the extra area is
OFF.
The position in the print area (20 cm) where each waveform is to be recorded is set
with Upper and Lower (%). The Upper value can be set in the range of 2% to 100%.
The Lower value can be set in the range of 0% to 98%. Each value can be set in 1%
steps. The minimum width is 2%.
[Example]
• When Flexible Zone on CH1 is set to Upper = 20% and Lower = 0%
The CH1 waveform is printed in the area from the bottom to 4 cm of the print area.
• When Flexible Zone on CH2 is set to Upper = 80% and Lower = 20%
The CH2 waveform is printed in the area from 4 cm to 16 cm from the bottom of the
print area.
Graticule
• Type
Select the graticule type from
, OFF, or
.
The grid setting is linked with the DISPLAY menu > Graticule setting. However, if
OFF is selected, the DISPLAY menu is set to
.
• Dark/Light
Set the graticule Dark/Light setting to Light or Dark.
• Width of the Vertical Graticule: Width
Select how to set the vertical graticule width. For details on the format of the
graticule lines that are printed, see section 9.3.
1Div:
Graticule obtained by dividing the print zone into 10 areas
10mm:
mm graticule type
Details
• Print Interval of Scale Values: Scale
Select the interval for the scale printed at the top and bottom edges of the print area
from OFF, 200 mm, 400 mm, or 800 mm. If OFF is selected, the scale is not printed.
• Print Interval of Channel Labels: CH Label
Select the interval for printing waveform channel labels from OFF, 200 mm, 400
mm, or 800 mm. The channel label is printed near each waveform. If OFF is
selected, the channel labels are not printed.
• Time Print: Time
Select the interval for printing the time from OFF, 200 mm, 400 mm, or 800 mm.
The time is printed at the top section of the print area. If OFF is selected, the times
are not printed.
• Gauge Print
Select whether to not print (OFF), print (ON), or print in detail (Fine) on the left side
of the print area. If ON is selected a gauge that equally divides the waveform
display area into two is displayed. If Fine is selected, a gauge that equally divides
the waveform display area into 10 is displayed. The gauge, scale values, an arrow
indicating the ground position, and V/div (Value/div)* are printed.
* Depending on the setting, V/div (Value/div) may not be printed.
12.1 Printing on the Built-in Printer