Strain Module (DSUB, Shunt-Cal) (701271) Specifications
Item
Specifications
Standard operating
Temperature: 23
±
5
°
C
conditions
Humidity:
20 to 80%RH
After a 30-minute warm-up and after calibration and auto balance
Effective measurement
–FS to +FS (set using upper and lower limits)
range
Number of input channels
2
Maximum sample rate
100 kS/s
Input format
DC bridge (auto balancing), balanced differential input, and isolated
Auto balance type
Electronic auto balance
Auto balance range
±
10000
µ
STR (1 gauge method)
Bridge voltage
Select from 2 V, 5 V, and 10 V.
Gauge resistance
120 to 1000
Ω
(bridge voltage: 2 V)
350 to 1000
Ω
(bridge voltage: 2 V, 5 V, and 10 V)
Gauge factor
1.90 to 2.20 (set in 0.01 steps)
Frequency characteristics
1
(–3 dB point when a sine wave with an amplitude equivalent to 60% of the range is applied) DC to
20 kHz
mV/V range support
Supports the strain gauge transducer unit system.
mV/V range = 0.5
×
(
µ
STR range/1000)
Measurement range (FS) and measurement range
When using STR range
Measurement Range (FS)
Measurement Range
500
µ
STR
–500 to +500
µ
STR
1000
µ
STR
–1000 to +1000
µ
STR
2000
µ
STR
–2000 to +2000
µ
STR
5000
µ
STR
–5000 to +5000
µ
STR
10000
µ
STR
–10000 to +10000
µ
STR
20000
µ
STR
–20000 to +20000
µ
STR
When using mV/V range
Measurement Range (FS)
Measurement Range
0.25 mV/V
–0.25 to +0.25 mV/V
0.5 mV/V
–0.5 to +0.5 mV/V
1 mV/V
–1 to +1 mV/V
2.5 mV/V
–2.5 to +2.5 mV/V
5 mV/V
–5 to +5 mV/V
10 mV/V
–10 to +10 mV/V
DC accuracy
1
±
(0.5% of FS + 5
µ
STR)
Maximum input voltage
Between Input+ and Input– 10 V (DC+ACpeak)
(At 1 kHz or less)
Maximum allowable
Between each terminal and earth ground
common mode voltage
42 V (DC+ACpeak) (CAT II, 30 Vrms)
(At 1 kHz or less)
Input connector
9-pin D-Sub connector (female)
Common mode rejection
80 dB (50/60 Hz) or more (typical
2
)
ratio
A/D conversion resolution
16 bit (48000 LSB/
±
FS: when Upper is +FS and Lower is –FS)
Temperature coefficient
Zero point:
±
5
µ
STR/
°
C (typical
2
)
Gain:
±
(0.02% of FS)/
°
C (typical
2
)
Bandwidth limit
Select from OFF, 1 kHz, 100 Hz, and 10 Hz
Cutoff characteristics: –12 dB/OCT (typical
2
)
Function
mV/V support. Supports the strain gauge transducer unit system.
Shunt calibration support. Built-in shunt calibration relay (1 gauge method).
Standard accessories
Connector shell set for soldering
A1520JD (9-pin D-Sub): 2 pieces, A1618JD (connector shell): 2 pieces
Compatible accessories (sold separately)
Recommended bridge head 701957 (D-Sub 120
Ω
, shunt-Cal, comes with a 5-m cable)
Recommended bridge head 701958 (D-Sub 350
Ω
, shunt-Cal, comes with a 5-m cable)
18.12 Module Specifications