Horizontal and Vertical Axes
5.15 Setting the Strain Measurement
<For a description of this function, refer to page 2-12.>
Procedure
Selecting the Channel
1.
Press
CH
.
2.
Press one of the soft keys from
CH1
to
CH16
in which the 701270
(STRAIN_NDIS) or 701271 (STRAIN_DSUB) is installed.
For the displayed menus when this operation is executed, see section 5.1.
Selecting the Measurement Range
3.
Press the
Range Unit
soft key, and then press the
µ
STR
(the unit of strain) or
mV/V
(the unit of the output value of the strain gauge transducer) soft key.
Note
Be sure to execute balancing when you change the measurement range.
If
µ
STR Is Selected
4.
Turn the
jog shuttle
to select the Range from 500
µ
STR to 20000
µ
STR.
If mV/V Is Selected
4.
Turn the
jog shuttle
to select the Range from 0.25 mV/V to 10 mV/V.
Note
If the range unit is set to mV/V, a numeric value is displayed at the right side of the range
display. This value is the maximum input at the current bridge voltage converted to volts. It
indicates the maximum input voltage at the selected range.
Setting the Display Range
5.
Press the
Upper/Lower
soft key to set the jog shuttle control to Upper.
6.
Turn the
jog shuttle
to set the upper limit.
7.
Likewise, set Lower.
Note
Pressing RESET sets the maximum and minimum values of the measurement range to the
upper and lower limits.
When Range
Unit is
µ
STR
When Range
Unit is mV/V