Item
Specifications
Immunity
Complying
EN61326-1 Table 2 (for use in industrial locations)
standard
EN61326-2-1
Already certified (701240/701250/701251/701255/701267/701261/701262/701265/701270/701271/
701275/701280/700986/700987/701955/701956/701957/701958)
• Influence in the immunity environment (performance criterion A)
Noise
701250:
≤
±
3 mV (1:1 input, 50 mV range conversion)
increase
701251:
≤
±
3 mV (1:1 input, 10 mV range conversion)
701255:
≤
±
25 mV (1:1 input, 50 mV range conversion)
701267:
≤
±
30 mV (1:1 input, 200 mV range conversion)
701261/701262:
≤
±
3 mV
701265:
≤
±
0.05 mV
701270:
≤
±
100
µ
STR
(when equivalent to
±
100 mV, gauge factor = 2, and bridge voltage = 2 V)
701271:
≤
±
100
µ
STR
(when equivalent to
±
100 mV, gauge factor = 2, and bridge voltage = 2 V)
701275:
≤
±
3 mV (1:1 input, 50 mV range conversion)
701280:
Within the specifications
Test conditions 701250:
10 MS/s, envelope mode, 500 mV range, no input filter,with the tip of the probe
(700929 (10:1)) shorted
701251:
1 MS/s, envelope mode, 100 mV range, no input filter,
with the tip of the probe (700929 (10:1)) shorted
701255:
10 MS/s, envelope mode, 500 mV range, no input filter,
with the tip of the probe (701940 (10:1)) shorted
701267:
100 kS/s, envelope mode, 200 mV range, no input filter,
with the tip of the cable
701261/701262: 100 kS/s, envelope mode, 50 mV range, no input filter,
With the end of the 3-m twisted-pair cable shorted
701265:
100 kS/s, envelope mode, 1 mV range, no input filter,
With the end of the 3-m twisted-pair cable shorted
701270:
100 kS/s, envelope mode, 500
µ
STR,
gauge factor: 2.0, no input filter,
701955 bridge voltage: 2 V
701956 bridge voltage: 10 V
701271:
100 kS/s, envelope mode, 500
µ
STR,
gauge factor: 2.0, no input filter,
701957 bridge voltage: 2 V
701958 bridge voltage: 10 V
701275:
100 kS/s, envelope mode, 500 mV range, no input filter, with the tip of the probe
(700929 (10:1)) shorted
701280:
25 kS/s, envelope mode, Frequency, 1 Hz range, no input filter, User (
±
1 V, threshold
level: 0 V, hysteresis:
±
1%), with the tip of the probe (700929 (10:1)) shorted
Test item
1. Electrostatic discharge
EN61000-4-2
Air discharge:
±
8 kV, contact discharge:
±
4 kV, criteria B
2. Radiated immunity
EN61000-4-3
80 MHz to 1 GHz, 1.4 GHz to 2 GHz, 10 V/m, criteria A
3. Conducted immunity
EN61000-4-6
3 V, criteria A
4. Electrical fast transient/burst
EN61000-4-4
Power line:
±
2 kV, signal line:
±
1 kV, criteria B
5. Power frequency magnetic field EN61000-4-8
30 A/m, 50 Hz, criteria A
6. Surge immunity
EN61000-4-5
Between lines:
±
1 kV, common:
±
2 kV, criteria B
7. Voltage dip and interruption
EN61000-4-11
0.5 cycle, both polarities, 100%, criteria A
Definitions of criteria • Criteria A Aforementioned “Influence in the immunity environment” is met during the test.
A and B
• Criteria B This apparatus continues to operate without hang-up or falling into uncontrollable
conditions during the test. No change of actual operating state or stored data is
allowed.
1 If the 701260 module is inserted into the SL1400, it will not comply with safety standard EN61010-1:2010 or EN61010-2-
030:2010.
2 The Overvoltage Category (Installation Category) is a value used to define the transient overvoltage condition and includes
the impulse withstand voltage regulation. I applies to electrical equipment that is powered by a circuit with overvoltage
control. II applies to electrical equipment that is powered by a fixed installation such as a distribution board.
18.11 General Specifications