18.1 Input Section
Item
Specifications
Number of input channels
16 ch 16-bit log (8 bits
×
2)
Type
Plug-in input unit (two 26-pin connectors (Logic A/Logic B) for the logic input)
Number of slots
8 (2 channels per slot)
Maximum record length
2.5 MW/CH (16 analog ch 16-bit logic),50 MW/channel max.
18.2 Trigger Section
Item
Specifications
Trigger mode
Auto, auto-level, normal, single, single(N), log, repeat (only in Chart Recorder mode)
Trigger level range
±
10 divisions around 0
Trigger hysteresis
When measuring voltage:
Select
±
0.1 div,
±
0.5 div, or
±
1 div of the trigger level
When measuring temperature:
Select
±
0.5
°
C,
±
1.0
°
C, and
±
2.0
°
C.
When measuring strain:
Select
±
2.5%,
±
12.5%, or
±
25% of the range
Trigger position
Can be set in 0.1% increments of the display record length.
Trigger delay range
0 to 10 s (resolution is 100 ns)
Hold off time range
0 to 10 s (resolution is 100 ns)
Manual trigger key
Dedicated manual trigger key is available
Simple trigger
Trigger source
CH1 to CH16, EXT (signal input from the TRIG IN terminal), LINE (commercial power supply signal
that is connected), Logic A, Logic B, and Time
Trigger slope
CH1 to CH16:
Rising, falling, or rising/falling
EXT, Logic A, Logic B: Rising or falling
Time trigger
Date (year/month/day), time (hour/minute), time interval (1 minute to 24 hours)
Enhanced trigger
Trigger source
CH1 to CH16, Logic A, and Logic B (AND and OR possible on each logic bit)
Trigger type
A->B(N):
Trigger occurs N
th
time condition B becomes true after condition A becomes true.
Count:
1 to 255
Condition A:
Enter/Exit
Condition B:
Enter/Exit
A Delay B:
Trigger occurs the first time condition B becomes true after the specified delay
elapses after condition A becomes true.
Delay:
0 to 10 s (resolution is 100 ns)
Condition A:
Enter/Exit
Condition B:
Enter/Exit
Chapter 18
Specifications