IM 701240-01E
5-8
5.5
Selecting the Input Coupling
<For a description of this function, refer to page 2-8.>
Procedure
Selecting the Channel
1.
Press
CH
.
2.
Press one of the soft keys from
CH1 to CH16
.
For the displayed menus when this operation is executed, see section 5.1.
Selecting the Input Coupling
3.
Press the
Coupling
soft key.
4.
Press the soft key corresponding to the desired input coupling.
•
To measure RMS values using the 701267 (HV (with RMS)), select AC-RMS or DC-
RMS. For the setup procedure of RMS measurements, see section 5.13.
•
To measure the temperature using the 701261 (UNIVERSAL), 701262 (UNIVERSAL
(AAF)), or 701265 (TEMP/HPV), select TC. For the setup procedure of temperature
measurements, see section 5.14.
•
To measure acceleration on the 701275 (ACCL/VOLT), select ACCL. For the setup
procedure of acceleration measurements, see section 5.16.
Select the input coupling
Explanation
Input Coupling
Select how the input signal is coupled to the vertical control circuit from the list below.
The selectable settings vary depending on the module.
AC
Acquires and displays only the AC component of the input signal.
DC
Acquires and displays all the components (DC and AC) of the input signal (selectable only when
measuring voltage).
GND
Checks the ground level.
TC
Select this type to measure temperature.
AC-RMS
Converts and displays only the AC component of the input signal.
DC-RMS
Converts and displays both DC and AC components of the input signal.
ACCL
Select this type to measure acceleration.