10-12
IM 701240-01E
Setting the Unit
14.
Use
jog SELECT
to set the Unit using up to four characters according
to the procedure given in section 4.2.
Setting the Label
15.
Use
jog SELECT
to set the Label according to the procedure given in
section 4.2. For a description of the Label display, see section 8.10, “Setting
Waveform Labels.”
Turning ON/OFF the Math Waveform Display
16.
Use
jog SELECT
to set Display to ON or OFF.
17.
Press
ESC
twice.
Selecting the Time (or the Number of Data Points) for Shifting the Phase
18.
Press the
Phase Shift
soft key. The phase dialog box opens.
19.
Use
jog SELECT
to set the time (or number of data points) for shifting
the phase of each waveform. If the time base is set to internal clock (Int), set the
time for shifting the phase. If the time base is set to external clock (Ext), set the
number of data points for shifting the phase.
ESC ESC
If the time base is set to
internal clock
If the time base is set to external clock, the setup
menu used to set the number of data points for
shifting the phase appears.
As necessary, set Math2 to Math8 in a similar fashion.
Executing the Computation
20.
Press the
Exec
soft key.