4-19
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Common Operations
Settings after Auto Setup
Waveform acquisition and display conditions
START/STOP
START
Acquisition mode
Normal
Number of acquisitions
Infinite
Time base
Int
Accumulate mode
OFF
Zoom target
Channels whose display is turned ON
Vertical-axis settings
Voltage range
Value that causes the absolute value of the input waveform
to be 1.6 to 4.5 divisions
Offset voltage
0 V or the center value between the maximum and
minimum values of the input signal
Coupling
DC
Bandwidth limit
FULL
Display ON/OFF
Turns ON channels whose absolute value of the input
voltage is greater than or equal to 20 mV (1:1)
Position
0 division
Variable mode
OFF
V Zoom
×
1
Scaling
OFF
Horizontal-axis settings
Record time
Value that the displays 1.6 to 4 periods of the waveform
with the shortest period of the auto setup target waveforms
Trigger settings
Trigger mode
Auto
Trigger type
SIMPLE
Trigger source
Channel whose waveform amplitude is greater than or
equal to 1 division with the longest period
Trigger level/slope
Center level between the maximum and minimum values/
rising
Trigger coupling
DC
Hysteresis
(high sensitivity)
Hold off time
0.0 ns
Trigger position
50.0%
Trigger delay
0.0
µ
s
Computation settings
Math
OFF
Other settings
Realtime output
OFF
Accumulation
OFF
Logic A and Logic B display
OFF
Items other than those listed above are not applicable for auto setup.
Note
If none of the input waveforms are detected on the target module, the waveform acquisition is
not started.