7-8
IM 701240-01E
7.4
Acquiring Data Using the Sequential Store
Function (Single (N) Mode)
<For a description of this function, refer to page 2-29.>
Procedure
1.
Press
MENU
.
2.
Press the
TRIGGER
soft key.
3.
Press the
Mode
soft key.
4.
Press the
Single(N)
soft key to set the trigger mode to Single (N).
5.
Press the
Single(N) Count
soft key.
6.
Turn the
jog shuttle
to set Single(N) Count.
Explanation
You can use the sequential store function by setting the trigger mode to Single(N).
Acquisition Count (Single(N) Count)
Set a value between 1 and 2000.
However, the selectable acquisition count varies depending on the selected record
length and acquisition mode. For details, see appendix 3.
Displaying Waveforms
You can recall waveforms from memory in the same fashion as with the history memory
function. For details, see section 11.1.
Notes on When Using Sequential Store
If you stop the waveform acquisition by pressing START/STOP, sequential store also
stops. Sequential store restarts from the beginning when acquisition resumes.