Preparations for Measurements
Connecting the Electromagnetic Pickup
• The SL1400 allows power-generating electromagnetic pickup to be connected
directly. The SL1400 does not support electromagnetic pickups that require external
power supply or those that require a terminator at the output.
• To connect electromagnetic pickups, use BNC cables. Use cables that are
appropriate for the electromagnetic pickups being used.
• When the input is set to EM Pickup (electromagnetic pickup), determination is not
made on whether the input voltage level exceeds the specified input voltage range.
The LEDs (see page 3-7) do not illuminate even when the voltage level is over range.
Power-generating
electromagnetic
pickup
Rotating object
to be measured
BNC connector
701280(FREQ)
SL1400
3.14 Connecting Sensors to the Frequency Module