Troubleshooting, Maintenance, and Inspection
Errors in Setting (800 to 899)
Code
Message and Corrective Action
Section
800
Illegal date-time. Set the correct date and time.
3.6
801
Illegal file name. The file name contains characters which are not allowed or the file name is not
Chapter 13
a valid MS-DOS file name. Enter another file name.
803
Turn off accumulate mode to enter history menu. It is not possible to activate the history menu
8.4
by pressing the HISTORY key during accumulation.
804
Cannot change this parameter while running. Press the START/STOP key to stop acquisition.
7.1
808
Can not be specified at the current record time setting.
5.2,
Appendix 1
809
Cannot change edge for status other than X.
Chapter 6
Set the state of the channel corresponding to condition A to ‘X’.
811
Cannot change display points in roll mode.
5.2
813
Set Items in measure menu. Set appropriate measurement items.
11.5
814
Duplicated Name. Change the label string.
8.10
816
Cannot turn off with the current record length. Please Change record length.
5.2
817
Cannot change. Please change X Trace in the X-Y menu.
8.6
819
Cannot change when Channel Display is OFF or Math settings are invalid.
5.1,
Set the channel display ON or make appropriate Math settings.
Chapter 10
821
Cannot change when External Clock is active.
5.20
823
Cannot change while running. Set the acquisition mode to Normal.
7.1, 7.2
824
Cannot change with the current acq mode.
7.2
829
Cannot change when Logic Mode is OFF or all bits of Logic Display are OFF.
5.18
834
Duplicate SCSI ID. Set different ID numbers.
13.7
835
Cannot change settings while realtime printing or realtime recording.
7.5
Stop printing or recording.
836
Cannot change settings during Action On Trigger. Stop the Action On Trigger.
6.18
837
Cannot set the channels which do not have modules installed.
–
838
Cannot set the channels which do not have the strain module installed.
–
839
Cannot Set or Execute.
–
840
If the trigger mode is set to Single, Single (N), or LOG, the acquisition mode cannot be set to
6.1, 7.2
Average.
843
If the acquisition mode is Average, the trigger mode cannot be set to Single, Single (N), or Log.
6.1, 7.2
844
It can’t be set at the time of the roll mode.
5.2
845
The trigger mode cannot be set to Single(N) during the roll mode.
5.2, 6.1
850
The acquisition mode cannot be set in the current record length.
5.2,
Appendix 1
851
Computation cannot be carried out at the current record length.
5.2, 10.3
855
Cannot be configured or executed during the history search operation.
11.2, 11.3
856
The record cannot be selected.
11.1 to 11.3
857
History record does not exist.
11.1 to 11.3
858
Cannot be configured or executed while computation is in progress.
11.1 to 11.3
Aborted when history display mode is set to One.
859
It can’t be set up or be carried out in the Preview mode.
12.1
Choose Quit, and cancel the Preview mode.
860
Cannot be configured or executed while updating the history all display.
11.1 to 11.3
Aborted when history display mode is set to One.
861
This format cannot output with color.
12.2, 12.3
868
Processing statistics.
11.6
To perform other operations, abort the statistical processing.
869
The channel which couldn’t be set up was specified.
Chapter 11,
13.8
870
Cannot be set when the acquisition mode is set to average.
7.2
872
Cannot be set during realtime recording.
7.5