Using the right or left arrow will scroll through the other setup parameters.
T E M P
L O W 1
# # .
# ° C
T E M P
L O W 2
# # .
# ° C
The TEMP LOW1 and TEMP LOW 2 behave similarly to the HIGH1 and 2 alarms.
This temperature setting adjacent to each parameter will trigger an alarm when the
bubbler temperature measurement is equal to or less than this temperature limit.
TEMP LOW1 has a corresponding relay output for remote annunciation. TEMP
LOW2 does not. TEMP LOW2 can be a higher, equal or lower temperature value
that TEMP LOW1.
5.7.3
Level Limit Alarms
L E V E L
4
# # .
#
L E V E L
3
# # .
#
L E V E L
2
# # .
#
L E V E L
1
# # .
#
L O W
C H E M
A L M
2 3
S P A R E
0
The LEVEL 1 through LEVEL 4 limit screen allows the user to define the level at which
the VaporGuard® will trip a given relay. However, these outputs are not alarms and
will not show up on the display nor will they be associated with the “ANY” ALARM.