The user may select between Quartz, SS Ultrasonic or SS No Level. For SS No Level
the user must remove the hardware interlock on JP4 of the Source Cup Assembly PCA
or the VaporGuard® will not function. (S.S. Optical Level is not available).
5.6.19
Product Selection
P R O D U C T
S E L E C T
X = E X I
T
√ = E X E C U T E
The user may select between;
Regular ( Trans-LC™ with maximum set point = 22 ºC)
POCl
3
(maximum set point = 26ºC)
Cable Code ( Trans-LC™ with maximum set point = 20ºC)
These are quartz bubbler applications and use the lowest thermal cutout setting (see
Section 5.9). The interlock should be used for all Class1, Division 2 applications. When
Cable Code is selected, the cable code alarm is activated.
5.6.20
Alarm Interlock
Interlocking an individual alarm will result in the removal of power to the heater/cooler
circuit. When an alarm is interlocked, it will automatically be enabled in the Alarm