5.3.3
Remote Temperature Output
This output is a 0 to +5 VDC signal with linear correspondence to a bubbler
temperature on a full dynamic range of 0 °C to 110 °C. Maximum bubbler
temperature will vary based on VaporGuard® model type but corresponding analog
output for all models will remain the same scale.
Likewise, +1 VDC output corresponds to a temperature reading of 22°C independent
of the maximum control temperature of each model.
5.3.4
Remote Display Module
The Remote Display Module allows the user to control, configure and monitor the
VaporGuard® from a remote location up to 7.6 meters (25 feet) away. The main
display on the Source Cup Assembly will also display the same information at the
same time.
However, the Remote Display Module can only be enabled to control and configure the
VaporGuard® once it has been enabled from the main display. Ensure to never install
temperature probe into this communications connector (J3).