Operating Instructions
Teledyne ML – T101 Operation Manual
72
NOTE
On analyzers with the multigas option activated (see Section s 4.5.1 and 6.8) the concentration value will
switch back and forth between from “H2S=XXX.X” to “SO2=XXX.X” depending on which gas is currently
being measured.
Also, The analyzer will switch between the HIGH and LOW analog reporting ranges
whenever the concentration level of the gas being currently measured fulfills the trigger
criteria listed at the beginning of this section.
4.4.4.7.
Range Units
The T101 can display concentrations in parts per billion (10
9
mols per mol,
PPB
), parts
per million (10
6
mols per mol,
PPM
), micrograms per cubic meter (µg/m
3
,
UG
) or
milligrams per cubic meter (mg/m
3
,
MG
). Changing units affects all of the display,
analog outputs, COM port and DAS values for all reporting ranges regardless of the
analyzer’s range mode.
To change the concentration units:
SETUP X.X
CONC UNITS: PPM
PPM PPB
UGM
MGM %
ENTER EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
CONC UNITS: PPB
PPM
PPB
UGM
MGM
ENTER
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
Select the preferred
concentration unit.
EXIT returns
to the main menu.
ENTR
accepts
the new unit,
EXIT
returns
to the
SETUP
menu.
NOTE
Concentrations displayed in mg/m
3
and µg/m
3
use standard temperature and pressure
(STP). The conversion factors from volumetric to mass units used in the T101 are:
SO
2
: ppb x 2.86 = µg/m
3
; ppm x 2.86 = mg/m
3
H
2
S: ppb x 1.52 = µg/m
3
; ppm x 1.52 = mg/m
3
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...