Operating Instructions
Teledyne ML – T101 Operation Manual
102
4.6.9.
TEST CHANNEL OUTPUT
When activated, output channel
A4
can be used to report one of the test functions
viewable from the SAMPLE mode display. To activate the A4 channel and select a test
function, follow this button sequence:
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
DIAG
ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG TCHN
TEST CHANNEL: NONE
NEXT
ENTR
EXIT
DIAG
TEST CHAN OUTPUT
PREV
NEXT
ENTR
EXIT
EXIT
returns
to the main
SAMPLE
display
Continue to press
NEXT
until …
DIAG TCHN
TEST CHANNEL: PMT READING
PREV
NEXT
ENTR
EXIT
Press
PREV
or
NEXT
to move through the
list of available
parameters
(Table 6-13)
Press
EXIT
to
return to the
DIAG
menu
Press
ENTR
to select
the displayed
parameter activating
the test channel.
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
Table 4-15. Test Parameters Available for Analog Output A4
TEST CHANNEL
TEST PARAMETER RANGE
1
NONE
Test channel is turned off
PMT READING
0-5000 mV
UV READING
0-5000 mV
SAMPLE PRESSURE
0-40 in-Hg-A
SAMPLE FLOW
0-1000 cm³/min
RCELL TEMP
0-70° C
CHASSIS TEMP
0-70° C
IZS TEMP
0-70° C
PMT TEMP
0-50° C
CHASSIS TEMP
0-70° C
HVPS VOLTAGE
0-5000 V
1
This refers to the voltage range of the parameter and
not the output signal of the test channel.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...