Teledyne ML - T101 Operation Manual
Principles Of Operation
257
10.3.2.
MULTIGAS MEASUREMENT & H
2
S
SO
2
SWITCHING
VALVE
When activated for operation the multigas measurement mode allows the instrument to
measure either or both H
2
S or SO
2
via a Teflon
®
switching valve. This valve, under
CPU control via the I
2
C buss and the relay board, directs the sample gas stream either
through the SO
2
scrubber and H
2
S
SO
2
converter (H
2
S measurement mode) or
directly to the sample chamber bypassing the H
2
S
SO
2
converter, allowing the
analyzer to measure SO
2
. The cycle for this operation is
Table 10-1. T101 Multigas Valve Cycle-Phases
Gas Mode
H
2
S
SO
2
Valve
Status
Default
Time
Settings
Activity
H
2
S
Gas stream directed
through scrubber and
converter
0 – 3
minutes
Wait period. Ensures sample chamber has
been flushed of previous gas.
3 – 10 m
Analyzer measures florescence in sample
chamber
SO
2
Gas stream bypasses
through scrubber and
converter
0 – 3
minutes
Wait period (dwell time). Ensures sample
chamber has been flushed of previous gas.
3 – 10 m
Analyzer measures florescence in sample
chamber
Cycle repeats every ~20Minuites
The timing of the above cycle is set by two variables (see Appendix A-2),
MEASURE_PERIOD,
which sets the total dwell time for each gas mode, and
MEASURE_DELAY
which sets the wait period before the instrument begins making
measurements after the gas mode has been switch.
10.3.3.
FLOW RATE CONTROL
The Model T101 uses a special flow control assembly located in the exhaust vacuum
manifold (Figure 10-7) to maintain a constant flow rate of the sample gas through the
instrument. This assembly consists of:
•
a critical flow orifice
•
two o-rings: Located just before and after the critical flow orifice, the o-
rings seal the gap between the walls of assembly housing and the critical
flow orifice.
•
a spring: Applies mechanical force needed to form the seal between the
o-rings, the critical flow orifice and the assembly housing.
10.3.3.1.
Critical Flow Orifice
The most important component of this flow control assembly is the critical flow orifice.
Critical flow orifices are a remarkably simple way to regulate stable gas flow rates. They
operate without moving parts by taking advantage of the laws of fluid dynamics. By
restricting the flow of gas though the orifice, a pressure differential is created. This
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...