Calibration Procedures
Teledyne ML – T101 Operation Manual
184
6.9.2.
AUTOCAL OF INSTRUMENTS IN MULTIGAS
MEASUREMENT MODE
If the analyzer is being operated in multigas measurement mode, auto cal can only be
programmed to calibrate the instrument with one of the available gases.
NOTE
Make sure that the gas for which the AutoCal sequence has been programmed is being
supplied to the analyzer at the time the autocal program initiates.
If the wrong gas is being supplied to the instrument, the analyzer will be mis-calibrated.
If the instrument is in multigas measurement mode and the
SNGL
reporting range mode
is selected, the AutoCal sequence is programmed as described above. If the reporting
range of the T101 must is set for either
IND
or
AUTO.
the gas type and range to be
measured must be specified. This parameter appears at the end of the programming
sequences after the
CALIBRATE: ON/OFF
parameter is set. For example:
SETUP X.X
GAS-RANGE: H2S-HIGH
<SET
EDIT
EXIT
SETUP X.X
CALIBRATE: ON
<SET
SET>
EDIT
EXIT
Follow standard AutoCal programming process to this
point, then …
SETUP X.X
GAS TO CAL: H2S
SO2
H2S
ENT
R EXIT
SETUP X.X
Gas TO CAL: SO2
LOW HIGH
ENTR
EXIT
EXIT
returns
to the
SETUP
Menu
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Sequence
Delta Time
MODE Delta Days
SETUP X.X
RANGE TO CAL: LOW
LOW
HIGH
ENT
R EXIT
SETUP X.X
RANGE TO CAL: HIGH
LOW HIGH
ENTR
EXIT
SETUP X.X
GAS-RANGE: SO2-LOW
<SET
EDIT
EXIT
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...