Teledyne ML – T101 Operation Manual
Operating Instructions
109
4.7.6.2.
Manually Configuring the Ethernet with Static IP Addresses
1. Connect a cable from the analyzer’s Ethernet port to a Local Area Network
(LAN) or Internet port.
2. Access Ethernet configuration through: SETUP>MORE>COMM>INET.
3. Follow the setup sequence as shown in the illustrations that follow, and
edit the Instrument and Gateway IP addresses and Subnet Mask to the
desired settings. (From the computer, enter the same information through
an application such as HyperTerminal).
Table 4-17. LAN/Internet Default Configuration Properties
PROPERTY
DEFAULT STATE
DESCRIPTION
DHCP
ON
This displays whether the DHCP is turned ON or
OFF. Press EDIT and toggle ON for automatic
configuration after first consulting network
administrator.
INSTRUMENT IP
ADDRESS
0.0.0.0
This string of four packets of 1 to 3 numbers each
(e.g. 192.168.76.55.) is the address of the
analyzer itself.
GATEWAY IP
ADDRESS
0.0.0.0
Can only be edited when DHCP is set to OFF.
A string of numbers very similar to the Instrument
IP address (e.g. 192.168.76.1.) that is the address
of the computer used by your LAN to access the
Internet.
SUBNET MASK
0.0.0.0
Can only be edited when DHCP is set to OFF.
Also a string of four packets of 1 to 3 numbers each
(e.g. 255.255.252.0) that identifies the LAN to
which the device is connected.
All addressable devices and computers on a LAN
must have the same subnet mask. Any
transmissions sent to devices with different subnets
are assumed to be outside of the LAN and are
routed through the gateway computer onto the
Internet.
TCP PORT
1
3000
This number defines the terminal control port by
which the instrument is addressed by terminal
emulation software, such as Internet or Teledyne
ML’s APICOM.
HOST NAME
T101
The name by which your analyzer will appear when
addressed from other computers on the LAN or via
the Internet. To change, see Section 4.7.6.3.
1
Do not change the setting for this property unless instructed to by Teledyne ML’s Technical
Support personnel.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...