Calibration Procedures
Teledyne ML – T101 Operation Manual
168
STEP THREE
: Perform the zero/span calibration:
M-P CAL
H2S STB=X.XXX PPB
H2S
=XXX.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
H2S STB=X.XXX PPB
H2S
=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
H2S STB=X.XXX PPB
H2S
=XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
H2S STB=X.XXX PPB
H2S
=XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL
RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > ENTR CONC
EXIT
Wait until
H2S STB
falls below
0.5 ppb.
This may take several
minutes.
The
SPAN
button now
appears during the
transition from zero to span.
You may see both buttons.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST
TST
>
CAL
SETUP
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
H2S
STB
test function.
This function calculates the
stability of the H
2
S
measurement
Press
ENTR
to changes the
OFFSET
&
SLOPE
values for the
H
2
S measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
H2S STB
may jump
significantly.
Wait until it falls back
below
0.5 ppb.
This may take several
minutes.
Press
ENTR to
change the
OFFSET
&
SLOPE
values for the
H
2
S measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
NOTE
If the ZERO or SPAN buttons are not displayed during zero or span calibration, the
measured concentration value is too different from the expected value and the analyzer
does not allow zeroing or spanning the instrument.
Consult Section 9.3 for more information on calibration problems.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...