SAMPLE PERIOD
is only used when the DAS parameter’s sample mode is set for
AVG, MIN
or
MAX
.
REPORT PERIOD
: Sets how often the sample readings stored in volatile memory are
processed, (e.g. average, minimum or maximum are calculated) and the results stored
permanently in the instrument’s Disk-on-module as well as transmitted via the analyzer’s
communication ports. The
REPORT PERIOD
may be set from the front panel.
If the
INST
sample mode is selected the instrument stores and reports an instantaneous
reading of the selected parameter at the end of the chosen
REPORT PERIOD
In
AVG, MIN
or
MAX
sample modes, the settings for the
SAMPLE PERIOD
and the
REPORT PERIOD
determine the number of data points used each time the average,
minimum or maximum is calculated, stored and reported to the COMM ports. The actual
sample readings are not stored past the end of the of the chosen
REPORT PERIOD
.
Also, the
SAMPLE PERIOD
and
REPORT PERIOD
intervals are synchronized to the
beginning and end of the appropriate interval of the instrument’s internal clock.
•
If SAMPLE PERIOD were set for one minute the first reading would occur
at the beginning of the next full minute according to the instrument’s
internal clock.
•
If the REPORT PERIOD were set for of one hour the first report activity
would occur at the beginning of the next full hour according to the
instrument’s internal clock.
EXAMPLE: Given the above settings, if DAS were activated at 7:57:35 the first sample
would occur at 7:58 and the first report would be calculated at 8:00 consisting of data
points for 7:58, 7:59, and 8:00.
During the next hour (from 8:01 to 9:00) the instrument will take a sample reading every
minute and include 60 sample readings.
When the
STORE NUM. SAMPLES
feature is turned on the instrument will also store
how many sample readings were used for the
AVG, MIN
or
MAX
calculation but not the
readings themselves.
REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF
If the instrument is powered off in the middle of a
REPORT PERIOD
, the samples
accumulated so far during that period are lost. Once the instrument is turned back on, the
DAS restarts taking samples and temporarily them in volatile memory as part of the
REPORT PERIOD
currently active at the time of restart. At the end of this
REPORT
PERIOD
only the sample readings taken since the instrument was turned back on will be
included in any
AVG
,
MIN
or
MAX
calculation. Also, the
STORE NUM. SAMPLES
feature will report the number of sample readings taken since the instrument was
restarted.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...