Troubleshooting & Service
Teledyne ML - T101 Operation Manual
242
QUESTION
ANSWER
How often do I need to
change the particulate
filter?
Once per week. Table 8-1 contains a maintenance schedule listing
the most important, regular maintenance tasks.
What is the averaging time
for a T101?
The default averaging time, optimized for ambient pollution
monitoring, is 240 seconds for stable concentrations and 20
seconds for rapidly changing concentrations; Refer to 10.7.1 for
more information.
My analyzer has the
optional, user -configurable
analog output channels.
How do I program and use
them?
Instructions for this can be found in the Manual Addendum for
Configurable Analog Output, PN 06270.
How long does the sample
pump last?
The sample pump should last about one year and the pump
diaphragms should to be replaced annually or when necessary.
Use the PRES test function displayed via the front panel to see if
the diaphragm needs replacement (refer to Section 9.1.2).
Do I need a strip chart
recorder or external data
logger?
No, the T101 is equipped with a very powerful internal data
acquisition system (DAS). Section 4.8 describes the setup and
operation in detail.
48.
9.8.
TECHNICAL ASSISTANCE
If this manual and its trouble-shooting / repair sections do not solve your problems,
technical assistance may be obtained from Teledyne ML, Technical Support, 9480
Carroll Park Drive, San Diego, CA 92121. Phone: +1 858 657 9800 or 1-800 324 5190.
Fax: +1 858 657 9816. Email: [email protected].
Before you contact Technical Support, fill out the problem report form in Appendix C,
which is also available online for electronic submission a
t http://www.teledyne-
ml.com/forms/index.asp.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...