Teledyne ML – T101 Operation Manual
Getting Started
53
3.6.
INITIAL CALIBRATION
3.6.1.
BASIC CALIBRATION PROCEDURE
The following three-step procedure assumes that the instrument does not have any of the
available zero/span (Z/S) or IZS valve options installed. Section 6 contains instructions
for calibrating instruments with valve options.
The initial calibration should be carried out with the analyzer’s reporting range for
SINGLE
range mode with a range span of
500 PPB
(factory default settings for most
units). This will enable you to compare your results to the factory calibration.
STEP ONE:
Set/verify the analog output reporting range of the T101:
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
Press this button to set
the analyzer for
SNGL
DUAL
or
AUTO
ranges
SETUP X.X
RANGE: 500.0 CONC
0 0
5
0 0 .
0
ENTR EXIT
To change the value of the reporting
range span, enter the number by
pressing the button under each digit
until the expected value appears.
Press this button to select the
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT
ignores the new setting and
returns to the
RANGE CONTROL
MENU
.
ENTR
accepts the new setting and
returns to the
RANGE CONTROL MENU
.
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
RANGE:
500.0 Conc
0 0 0 5 0 .0
ENTR
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...