
Operating Instructions
Teledyne ML – T101 Operation Manual
94
To adjust the zero and span values of the current outputs, activate the ANALOG I/O
CONFIGURATION MENU (see Section 4.6.3), then press
:
DIAG AIO
CONC_OUT_2 SPAN: 5000 mV
U100 UP10
UP DOWN DN10
D100
ENTR EXIT
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
DIAG AIO
AIN CALIBRATED:
NO
SET>
EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_2 RANGE: CURR
<SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2:CURR, NO CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the analog output channel
to be configured:.
Increase or decrease the current
output by 100, 10 or 1 counts.
The resulting change in output
voltage is displayed in the upper
line.
Continue adjustments until the
correct current is measured with
the current meter.
DIAG AIO
CONC_OUT_2 ZERO: 27 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
EXIT
ignores the
new setting,
ENTR
accepts the new
setting.
DIAG AIO
CONC_OUT_2 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 ZERO: 4921 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
DIAG AIO
CONC_OUT_2 CURR-Z: 0 mV
U100
UP10
UP DOWN DN10 D100
ENTR EXIT
The instrument attempt to automatically calibrate
the channel … then beep.
If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250
Ω
±
1% resistor across the current loop output. Using a
voltmeter, connected across the resistor, follow the procedure above but adjust the output
to the following values:
Table 4-14. Current Loop Output Calibration with Resistor
FULL
SCALE
VOLTAGE FOR 2-20 MA
(measured across resistor)
VOLTAGE FOR 4-20 MA
(measured across resistor)
0%
0.5 V
1.0 V
100%
5.0 V
5.0 V
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...