Teledyne ML – T101 Operation Manual
Operating Instructions
99
4.6.6.
LAMP CALIBRATION
An important factor in accurately determining H
2
S concentration, once the H
2
S is
converted to SO
2,
is the amount of UV light available to transform the SO
2
into SO
2
*
(see Sections 10.1.1 and 10.1.2). The Model T101 compensates for variations in the
intensity of the available UV light by adjusting the H
2
S concentration calculation using a
ratio (
LAMP RATIO
) that results from dividing the current UV lamp (
UV LAMP)
intensity by a value stored in the CPU’s memory (
LAMP_CAL
). Both LAMP Ratio and
UV Lamp are test functions viewable from the instrument’s front panel.
To cause the analyzer to measure and record a value for
LAMP_CAL
, press:
Exit at
any time
to return
to main
the
SETUP
menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing
NEXT
until . .
.
DIAG
LAMP CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG FCAL
LAMP CAL VALUE:4262.4 mV
4 2 6 2 .4 ENTR EXIT
E
NTR
accepts the
new value
EXIT
ignores the new
value
The value displayed is the
current output of the UV
source reference detector
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...