Teledyne ML - T101 Operation Manual
Troubleshooting & Service
229
does not turn on/off, then the associated device (valve or heater) or its control device
(valve driver, heater relay) is malfunctioning.
Several of the control devices are in sockets and can easily be replaced. The table below
lists the control device associated with a particular function:
Table 9-6. Relay Board Control Devices
FUNCTION
CONTROL DEVICE
SOCKETED
Valve0 – Valve3
U5
Yes
Valve4 – Valve7
U6
Yes
All heaters
K1-K5
Yes
9.5.9.
MOTHERBOARD
9.5.9.1.
A/D functions
A basic check of the analog to digital (A/D) converter operation on the motherboard is
to use the
Signal I/O
function under the
DIAG
menu. Check the following two A/D
reference voltages and input signals that can be easily measured with a voltmeter.
•
Using the Signal I/O function (Section 4.6.1 and Appendix D), view the
value of REF_4096_MV and REF_GND. If these signals are within 10
mV and 3 mV, respectively, of their nominal values (4096 and 0) and are
stable to within ±0.5 mV, the basic A/D converter is functioning properly.
If these values fluctuate largely or are off by more than specified above,
one or more of the analog circuits may be overloaded or the
motherboard may be faulty.
•
Choose one parameter in the Signal I/O function such as
SAMPLE_PRESSURE (see previous section on how to measure it).
Compare its actual voltage with the voltage displayed through the
SIGNAL I/O function. If the wiring is intact but there is a difference of
more than ±10 mV between the measured and displayed voltage, the
motherboard may be faulty.
9.5.9.2.
Analog Output Voltages
To verify that the analog outputs are working properly, connect a voltmeter to the output
in question and perform an analog output step test as described in Section 4.6.2.
For each of the steps, taking into account any offset that may have been programmed
into the channel (Section 4.6.3.4), the output should be within 1% of the nominal value
listed in the table below except for the 0% step, which should be within 2-3 mV. If one
or more of the steps is outside of this range, a failure of one or both D/A converters and
their associated circuitry on the motherboard is likely.
Table 9-7. Analog Output Test Function - Nominal Values
FULL SCALE OUTPUT VOLTAGE
100MV
1V
5V
10V
STEP
%
NOMINAL OUTPUT VOLTAGE
1
0
0 mV
0
0
0
2
20
20 mV
0.2
1
2
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...