Operating Instructions
Teledyne ML – T101 Operation Manual
116
After connecting a USB cable between your PC and the instrument, ensure their baud
rates match (change the baud rate setting for either your PC’s software or the instrument).
The baud rate setting is in the Communications Menu under COM2, which is the default
setup menu for USB configuration.
Also, while there are various communication modes available (Table 4-18), the default
settings are recommended for USB, except to change the baud rate if required..
4.7.8.
MULTIDROP RS-232 SET UP
When the RS-232 Multidrop option is installed, connection adjustments and configuration
through the menu system are required. This section provides instructions for the internal
connection adjustments, then for external connections, and ends with instructions for
menu-driven configuration.
NOTE
Because the RS-232 Multidrop option uses both the RS232 and COM2 DB9 connectors on the analyzer’s
rear panel to connect the chain of instruments, COM2 port is no longer available for separate RS-232 or
RS-485 operation.
CAUTION
Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges too small to be felt by the
human nervous system. Failure to use ESD protection when working with electronic assemblies will void
the instrument warranty. Refer to Section 13 for more information on preventing ESD damage.
In each instrument with the Multidrop option there is a shunt jumpering two pins on the
serial Multidrop and LVDS printed circuit assembly (PCA), as shown in Figure 4-9. This
shunt must be removed from all instruments except that designated as last in the
multidrop chain, which must remain terminated. This requires powering off and opening
each instrument and making the following adjustments:
1.
With NO power to the instrument, remove its top cover and lay the rear
panel open for access to the Multidrop/LVDS PCA, which is seated on the
CPU.
2.
On the Multidrop/LVDS PCA’s JP2 connector, remove the shunt that
jumpers Pins 21
↔
22 as indicated in Figure 4-9. (Do this for all but the
last instrument in the chain where the shunt should remain at Pins 21
↔
22).
3.
Check that the following cable connections are made in all instruments
(again refer to Figure 4-9):
•
J3 on the Multidrop/LVDS PCA to the CPU’s COM1 connector
(Note that the CPU’s COM2 connector is not used in Multidrop)
•
J4 on the Multidrop/LVDS PCA to J12 on the motherboard
•
J1 on the Multidrop/LVDS PCS to the front panel LCD
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...