Calibration Procedures
Teledyne ML – T101 Operation Manual
182
Toggle to
set
number of
days
between
procedures
(1-367)
Toggle to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP X.X
DELTA DAYS: 1
<SET SET>
EDIT
EXIT
SETUP X.X
DELTA DAYS: 1
0 0 2
ENTR
EXIT
SETUP X.X
DELTA DAYS:2
<SET
SET>
EDIT
EXIT
SETUP X.X
DELTA TIME00:00
<SET SET>
EDIT
EXIT
SETUP X.X
DELTA TIME: 00:00
0 0 :3 0
ENTR
EXIT
SETUP X.X
DELTA TIEM:00:30
<SET
SET>
EDIT
EXIT
SETUP X.X
STARTING TIME:14:15
<SET
SET>
EDIT
EXIT
EXIT
returns
to the
SETUP
Menu
SETUP X.X
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP X.X
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP X.X
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Toggle to
set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP X.X
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP X.X CALIBRATE: OFF
ON
ENTR
EXIT
SETUP X.X
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle
button
between
Off and
ON
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SEQ 1) DISABLED
NEXT
MODE
EXIT
SETUP X.X
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X
MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO–SPAN
PREV
NEXT
ENTR
EXIT
SETUP X.X
SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT
MODE
SET
EXIT
SETUP X.X
TIMER ENABLE: ON
SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Toggle to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
Default
value is
ON
SETUP X.X
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
SETUP X.X
STARTING TIME:00:00
<SET SET>
EDIT
EXIT
Toggle to set
time:
Format : HH:MM
This is a 24 hr
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL CALZ CALS
SETUP
With dynamic calibration turned on, the state of the internal setup variables dyn_Span and
DYN_ZERO
is set to
ON
and the instrument will reset the slope and offset values for the
H
2
S response each time the AutoCal program runs. This continuous re-adjustment of
calibration parameters can often mask subtle fault conditions in the analyzer.
Содержание T101
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank ...
Страница 10: ...vi This page intentionally left blank ...
Страница 12: ...viii This page intentionally left blank ...
Страница 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Страница 34: ...Getting Started Teledyne ML T101 Operation Manual 30 3 2 2 REAR PANEL Figure 3 4 Rear Panel Layout ...
Страница 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Страница 190: ...Calibration Procedures Teledyne ML T101 Operation Manual 186 This page intentionally left blank ...
Страница 202: ...Instrument Maintenance Teledyne ML T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank ...
Страница 204: ...Instrument Maintenance Teledyne ML T101 Operation Manual 200 This page intentionally left blank ...
Страница 247: ...Teledyne ML T101 Operation Manual Troubleshooting Service 243 This page intentionally left blank ...
Страница 248: ......
Страница 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Страница 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Страница 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Страница 345: ...D 7 07266B DCN6485 ...