10 Installation: Ceph arm
Installation Manual ORTHOPHOS XG 3D / Ceph
10.3 Installation
båÖäáëÜ
9. Use the arrow keys (I) on the control panel to lower the unit until the
bearing bolts on the unit are at the same height as the drillings on the
ceph support arm.
NOTICE! Check the room height before you raise the unit. If the room
height is less than 2.27m (89 3/8") or 2.30m (90 1/2") for installation
with floor stand, you must limit the maximum travel height (S018: Test
step 2).
Installing the ceph arm
1. Attach the support arm to the four bearing bolts.
NOTICE! The connecting cables must not be crushed. Make sure that
the connecting cables lie in the groove of the support arm and are not
crushed.
2. Screw in the conical nuts (J) and tighten them firmly.
Tip: The support arm may get tilted while you are tightening the
conical nuts. You can avoid this by lifting and moving the support arm
gently while you are tightening the nuts.
3. Remove the styrofoam packaging from the ceph.
4. Follow the instructions in the chapter entitled Installing the secondary
4x
J