13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1 General information about unit adjustment and calibration
båÖäáëÜ
13.1.6
Test phantoms for adjustment and calibration
13.1.6.1
OP XG needle phantom for panoramic adjustment
Warning for XG 5 / 3D / 3D ready ceph exposures
For panoramic sensor and pan symmetry panorama adjustments, the OP
XG needle phantom (A) must be inserted into the pan bite block holder of
the unit. For pan diaphragm panorama adjustment, the needle phantom
must be taken out of the bite block holder.
NOTICE
Risk of damage to unit
Make absolutely sure that the OP XG needle phantom is inserted in the
bite block holder of the unit in such a way that the front needle (B) of the
phantom points
away from the unit in the direction of the patient
(see
figure). Otherwise, this can result in a collision between the combisensor
and the OP XG needle phantom and thereby damage the unit.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the device
before
a ceph exposure is taken; otherwise, the
phantom could collide with the sensor.
A
B