13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
At the bottom of the action area, the two values (L)x and (L)y are
suggested for adjusting the ceph secondary diaphragm.
IMPORTANT! If the suggested values are greater than ±0.5 mm,
the diaphragm must be mechanically adjusted.
Positive sign =
Moves the diaphragm to the right or upward
Negative sign =
Moves the diaphragm to the left or downward
4. If necessary, adjust the diaphragm mechanically at this point [ → 261].
5. Confirm the mechanical correction of the ceph secondary diaphragm
by clicking on
"OK"
.
SIDEXIS XG is ready for exposure [ → 206] again.
6. Take an exposure (80 kV/14 mA; 0.6 s) [ → 206].
SIDEXIS XG calculates the adjustment values S1, S2, and S3
automatically from the exposure and enters them in the text
boxes of the
"Ceph - Fixed point of rotation"
menu.
The exposure is displayed in the exposure window in SIDEXIS
XG.