13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.4 Mechanical adjustments
Installation Manual ORTHOPHOS XG 3D / Ceph
3. Set the diaphragm tilt via screw (C) (Lx in mm) and the diaphragm
height via screw (D) (Ly in mm).
Screw (C):
Turn counterclockwise = adjusts the diaphragm to the right
Turn clockwise = adjusts the diaphragm to the left
Screw (D):
Turn counterclockwise = adjusts the diaphragm downward
Turn clockwise = adjusts the diaphragm upward
To measure the adjustment, use a gauge.
The above-mentioned directions of rotation for the screws apply to
both left-arm and right-arm versions.
4. Retighten screws (B) firmly.
5. Once adjustment is complete, put the secondary diaphragm cover
back again and fix it in place with the screw (A).