Sirona Dental Systems GmbH
13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1 General information about unit adjustment and calibration
båÖäáëÜ
13.1.6.2
OP XG 3D ceph adjustment phantom for the adjustment of the
cephalometer
For ceph diaphragm and ceph main X-ray beam direction 2D adjustment,
you must insert the OP XG ceph adjustment phantom (A) into the cover
of the dual sensor.
You must take the adjustment phantom back out again for ceph fixed
rotation point 2D adjustment.
A