13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2.2.2
Aperture
✔
The geometry phantom
is not
in the pan bite block holder on the unit.
1. In the structure tree, under
"3D Adjustment/calibration"
, click on the
"Diaphragm"
element (S030.10).
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (85 kV/0.33 mAs) [ → 206].
A message is displayed in the action window to indicate whether
adjustment has been successful and the traffic light symbol in
front of the
"Diaphragm"
element in the structure tree responds
accordingly:
Adjustment not OK = traffic light stays red
Adjustment OK = traffic light turns yellow
4. Repeat the procedure starting at Step 2 until adjustment is
successful.