13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
2. Once you have taken the exposure, measure distance S1 with the
SIDEXIS XG measuring ruler.
3. Overwrite the adjustment value for S1 in the text box of the
"Ceph -
Fixed point of rotation Quickshot"
menu with the measured value.
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area". Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Ceph - QuickShot fixed point of