11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.4.1
S018: Test step 2
Set the maximum travel height
S018_2 XG 3D / XG 3D ready
1. Move the unit to the required maximum travel height by pressing the
Up/Down keys in the user mode on the control panel.
2. Call service routine S018.2 [ → 125].
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.
3. To save the maximum travel height, touch the Memory key (B) (R key
(C) lights up) followed by the R key (C).
4. Set the mechanical limit stop at the unit:
Loosen nut (D) and move mechanical limit stop (E) for the limit switch
until it engages. Tighten the nut (D) again.
The next time the UP key is pressed, the unit stops 10 mm below the
limit switch.
5. Exit the service routine [ → 129].
D
E
*